Close

1. Identity statement
Reference TypeConference Paper (Conference Proceedings)
Sitemarte.sid.inpe.br
Holder Codeisadg {BR SPINPE} ibi 8JMKD3MGPCW/3DT298S
Repositorysid.inpe.br/deise/1999/02.01.11.04
Last Update2005:02.17.17.06.22 (UTC) administrator
Metadata Repositorysid.inpe.br/banon/2001/03.14.09.42.17
Metadata Last Update2018:06.06.02.48.26 (UTC) administrator
Secondary KeyINPE-6162-PRE/2251
ISBN85-17-00014-5
Label7526
Citation KeySilvaMoraMáxiViei:1996:MeAfEs
TitleMetodologia de aferição espectral para espectrorradiômetro
Format[CD-ROM]
Year1996
Secondary Date19970610
Access Date2024, May 19
Secondary TypePRE CN
Number of Files1
Size1065 KiB
2. Context
Author1 Silva, Davi Affonso da
2 Moraes, Elisabete Caria
3 Máximo, Orlando Alves
4 Vieira, Pedro Ronald
Resume Identifier1
2 8JMKD3MGP5W/3C9JH24
Group1
2 DSR-INPE-MCT-BR
Affiliation1 Centro Técnico Aeroespacial (CTA). Instituto de Estudos Avançado (IEAv).
2 Instituto Nacional de Pesquisas Espaciais (INPE)
3 Centro Técnico Aeroespacial (CTA). Instituto de Estudos Avançado (IEAv).
4 Instituto Nacional de Pesquisas Espaciais (INPE)
Author e-Mail Address1
2 bete@met.inpe.br
EditorKrug, Thelma
e-Mail Addresserich@sid.inpe.br
Conference NameSimpósio Brasileiro de Sensoriamento Remoto, 8 (SBSR).
Conference LocationSalvador
Date14-19 abr. 1996
PublisherINPE
Publisher CitySão José dos Campos
Pages957-964
Book TitleAnais
Tertiary TypeArtigos
OrganizationInstituto Nacional de Pesquisas Espaciais
History (UTC)2005-01-06 19:34:51 :: lise -> administrator ::
2009-06-30 13:55:35 :: administrator -> erich@sid.inpe.br ::
2010-05-14 02:53:36 :: erich@sid.inpe.br -> marciana ::
2011-02-16 12:53:09 :: marciana -> administrator :: 1996
2018-06-06 02:48:26 :: administrator -> :: 1996
3. Content and structure
Is the master or a copy?is the master
Content Stagecompleted
Transferable1
KeywordsSENSORES E TECNICAS DE SENSORIAMENTO REMOTO
ESPECTRORRADIOMETRO
RESOLUCAO ESPECTRAL
METODOLOGIA
SPECTRON SE-590
AbstractThis work describes a methodology to evaluate the spectral resolution of the spectroradiometer Spectron SE-590. The waveband checking of the SE-590 was performed by Material and Sensors Laboratory at INPE, using a grating monochromator model 7240. In the spectral range between 400 and 700nm the monochromator was ajusted for a waveband range of 6.2nm, while in the spectral range between 700 and 1100nm a waveband range of 12.4nm was used. Differences among spectral detector heads from the SE-590 were observed. Therefore, it is recomended to have a reference calibrated detector head in order to correct for errors in the BRF that appear while using uncalibrated spectral detector head(s)in field works.
AreaSRE
TypeRadiometria
Arrangementurlib.net > BDMCI > Fonds > Produção anterior à 2021 > DIDSR > Metodologia de aferição...
doc Directory Contentaccess
source Directory Content
T176.DOC 01/02/1999 10:05 328.0 KiB 
agreement Directory Contentthere are no files
4. Conditions of access and use
data URLhttp://urlib.net/ibi/sid.inpe.br/deise/1999/02.01.11.04
zipped data URLhttp://urlib.net/zip/sid.inpe.br/deise/1999/02.01.11.04
LanguagePt
Target FileT176.pdf
User Groupadministrator
erich@sid.inpe.br
Visibilityshown
Read Permissionallow from all
5. Allied materials
Mirror Repositorydpi.inpe.br/marte@80/2007/10.17.19.59
Next Higher Units8JMKD3MGPCW/3ER446E
Citing Item Listsid.inpe.br/mtc-m21/2012/07.13.14.45.11 1
Dissemination<E>
Host Collectiondpi.inpe.br/banon/2003/12.10.19.30
6. Notes
Empty Fieldsarchivingpolicy archivist callnumber contenttype copyholder copyright creatorhistory descriptionlevel documentstage doi edition identifier issn lineage mark nextedition notes numberofvolumes orcid parameterlist parentrepositories previousedition previouslowerunit progress project readergroup rightsholder schedulinginformation secondarymark serieseditor session shorttitle sponsor subject tertiarymark url versiontype volume


Close